FTIR FUNCTIONAL GROUP

C–Br stretching in FTIR: a weak band below 700 cm⁻¹

Bromine rarely gives an obvious band. The C–Br stretch is weak, low in frequency, and easy to confuse with C–Cl, C–I or a feature from the sample preparation itself, so it is identified from a cluster of evidence rather than from one number. Every assignment on this page is quoted from a published spectrum.

6 published spectra quoted below

Direct answer

Look for C–Br between roughly 500 and 700 cm⁻¹, most often in the 570–680 cm⁻¹ window, and expect a weak, sometimes broad band. The published spectra quoted here place it at 574.11, 601, 616.6, 655.80, 663, 675 and 677 cm⁻¹. A band in that window supports a bromide only when the rest of the spectrum and an elemental check agree.

What decides the assignment

  1. A heavy atom, so a low wavenumber

    Bromine is much heavier than carbon, and the C–Br stretch falls well below the C–Cl and C–F regions. A sharp band near 1200 cm⁻¹ is not C–Br, however convenient that would be for a clean report. When nothing appears below about 700 cm⁻¹, bromine is simply not visible in the mid-IR window you measured.

  2. The band is weak, and that is the main difficulty

    The C–Br bond has a small dipole change, so the stretch is typically a weak, sometimes broad feature rather than a headline peak. Yassin et al. describe the 574.11 cm⁻¹ band as weak and broad, and place a weak C–Cl band at 752.97 cm⁻¹ in the same spectrum. A missing C–Br band therefore does not prove that no bromine is present.

  3. Halogen bands overlap each other

    C–Cl, C–Br, C–I and several ring or skeleton modes share the same low-frequency window. Al-Otibi et al. assign 601 cm⁻¹ to "C–Br or C–I", which states the ambiguity honestly instead of resolving it. The neighbouring bands and the sample history decide which halogen reading is plausible.

  4. Preparation adds bands of its own

    Potassium bromide is both a common pellet matrix and a source of confusion in this region; Lefsih et al. assign a 690 cm⁻¹ peak to C–Br formed while preparing pellets in KBr. ATR crystals also lose throughput at the bottom of their range, which matters exactly where C–Br absorbs. Record the sampling mode with the result.

  5. The reported position scatters, so use a window

    Across the spectra quoted here the assigned C–Br position moves from 574.11 to 677 cm⁻¹ with the matrix and the neighbouring groups. Treat 570–680 cm⁻¹ as a search window and check shape and relative intensity, not a single textbook number.

Where the band falls

Range Group What it means here What to check
570–610 cm⁻¹ C–Br, the most frequently quoted window Most assignments in this page fall here: 574.11 cm⁻¹ in Yassin et al., 601 cm⁻¹ in Al-Otibi et al. and 616.6 cm⁻¹ in Kiran et al. Expect a weak band. Check that C–Cl, C–I or a matrix feature cannot explain it before naming the halogen.
610–660 cm⁻¹ C–Br in complex, hydrogen-bonded samples Bawazeer et al. place C–Br at 655.80 cm⁻¹ in a plant extract, and Jing et al. assign 663 cm⁻¹ to C–Br alongside a C–N band at 1025 cm⁻¹. C–N, C–O and C–S contributions crowd this region. Separate them with the 1000–1300 cm⁻¹ bands first.
660–695 cm⁻¹ Upper end of the reported range and the KBr pellet region Chowdhury et al. report C–Br at 675 and 677 cm⁻¹ in two formulations of the same starch film. The same window is where KBr pellet artefacts appear near 690 cm⁻¹. Re-run the sample by ATR or as a mull, or check the blank pellet, before calling a 690 cm⁻¹ band C–Br.
500–570 cm⁻¹ Lower-frequency end of the halide region Heavier bromide environments and some aryl bromides absorb here, close to the bottom of the range of many mid-IR instruments. Confirm the spectrometer and the ATR crystal still have usable throughput this low before interpreting a band there.

The ranges overlap on purpose: that is why a single position rarely settles the assignment.

Where this group sits in the reference spectra

Wavenumber (cm⁻¹) Supporting facts Cited sources
1545 6 5
1490 4 2
1637 2 2
1540 2 2
500 2 2
1400 2 2
651 2 2
1435 2 1
990 1 1
402 1 1
1680 1 1
1393 1 1
3400 1 1
3612 1 1
1539 1 1
1495 1 1
3630 1 1
1450 1 1
3408 1 1
2159 1 1
3375 1 1
2170 1 1
3610 1 1
961 1 1
1551 1 1
1662 1 1
540 1 1
1444 1 1
751 1 1
970 1 1

Showing the 30 best-supported peaks of 78 total for Bromine.

Possible materials

Material Supporting peaks Cited sources
ZrO2 1743, 455, 460 1

Check the bands around it

  • C–H stretching, 2850–2960 cm⁻¹

    An alkyl bromide needs an alkyl framework. Without C–H stretching, a band near 600 cm⁻¹ is more likely inorganic or an artefact than an alkyl bromide.

  • Aromatic C–H out-of-plane, 700–900 cm⁻¹

    For an aryl bromide the substitution pattern in this region usually says more than the C–Br stretch itself, because ring modes couple into the same window.

  • C=O stretching, 1700–1750 cm⁻¹

    Acyl bromides hydrolyse quickly. If a carbonyl band and a halide band appear together, consider a bromo ketone rather than a simple alkyl bromide.

  • C–Cl stretching, 700–800 cm⁻¹

    The C–Cl band usually sits above the C–Br band. Seeing both, as in the Yassin spectrum, makes a mixed halide sample more plausible than either assignment alone.

Published spectra

Expert reading

Treat every C–Br call as provisional and say so in the report. Give the position, the bandwidth and the sampling mode (KBr pellet, film or ATR), and list the bands that support a halogen. When the answer has consequences — a restricted-substance screen, a flame-retardant claim, a failed batch — send the sample for XRF or combustion ion chromatography instead of arguing from a weak band near 600 cm⁻¹. FTIR is good at telling you where to look and poor at telling you how much bromine is there.